Kern Moisture Analyser DAB 100-3 which is particularly easy to use – now also as a version with [d] = 10 mg, ideal for recurring rapid tests.
Product Description
Standard







- Backlit graphic display, digit height 14mm, digit height 14 mm
- Drying process active
- Active heating profile
- Active switch-off criteria
- Previous drying time
- Current temperature
- Current moisture content in %
- Halogen quartz glass heater 400 W
- Observation window above the sample, useful during initial setting
- Internal memory for automatic sequence of 15 complete drying processes and 5 drying processes carried out
- The last value measured remains on the display until it is replaced by a new measurement
- 50 sample plates included
- Application handbook: On the internet, you will find a practical application handbook containing many examples, field reports, settings and tips for each KERN moisture analyser
Technical Specification
Construction
| Level indicator | Yes |
|---|---|
| Material weighing plate | Aluminium dish |
| Overall dimensions mounted (W×D×H) | 240 mm x 365 mm x 180 mm |
| Revolving screw feet | Yes |
| Weighing surface (d) | 92 mm |
Services (optional)
| DAkkS Certificate | 963-127 |
|---|---|
| Factory calibration certificate (temperature) | 964-305 |
Display
| Display digit height (large) | 14mm |
|---|---|
| Display digit height (small) | 6mm |
Power supply
| Input voltage | 230 V AC 50 Hz |
|---|---|
| Mains adapter/adapter included | EURO |
Category
| Brand | KERN |
|---|---|
| Product Group | Moisture analyser |
Measuring system
| Adjustment options | External calibration |
|---|---|
| Eccentric loading at 1/3 [Max] | 5 mg |
| Linearity | 3 mg |
| Measuring capacity [Max] | 199 °C |
| Minimum weight (USP) | 2 g |
| Readability (moisture) | 0,010 % |
| Readability [d] | 1 mg |
| Repeatability | 1 mg |
| Resolution | 110000 |
| Stabilisation time under laboratory conditions | 3 s |
| Tare range | 110 g |
| Warm up time | 2 h |
| Weighing capacity [Max] | 110 g |
| Weighing system | Strain gauge |
| Weighing units | g % Feuchte % Trockenmasse |
Environmental conditions
| Maximum humidity | 80% |
|---|---|
| Maximum operating temperature | 40 °C |
| Minimum ambient temperature | 5 °C |
| Minimum humidity | 5 % |
Packaging
| Dimensions packaging (W×D×H) | 490 mm x 350 mm x 350 mm |
|---|---|
| Gross weight | 7,100 kg |
| Net weight | 4,820 kg |