Memmert

Memmert UF750plus Universal Oven

The Memmert UF750plus Universal Oven is a high-capacity forced convection laboratory oven designed for large-scale drying, heating, and testing applications. With its 749-litre chamber and advanced TwinDISPLAY control system, it delivers precise temperature control, uniform heat distribution, and full process monitoring for demanding laboratory and industrial environments.

Category: Tag: Brand:

Memmert UF750plus Universal Oven

Key Features of Memmert UF750plus

  • Large 749-litre capacity for high-volume applications
  • TwinDISPLAY ControlCOCKPIT with dual TFT screens
  • Forced convection system for excellent temperature uniformity
  • Precise temperature control up to 300°C
  • HeatBALANCE function for adjustable heat distribution
  • Dual Pt100 sensors for enhanced monitoring and accuracy
  • Integrated data logging with USB and Ethernet connectivity
  • AutoSAFETY system with multi-level overtemperature protection
  • Designed for continuous operation in demanding environments

Applications

The Memmert UF1060plus universal oven is suitable for advanced laboratory environments where precision and repeatability are critical:

  • Large-scale laboratory drying and heating
  • Industrial testing and batch processing
  • Pharmaceutical and regulated environments
  • Quality control and validation processes
  • Research requiring data logging and traceability

Why Choose Memmert Ovens

Memmert ovens are trusted worldwide for:

  • High precision and reliability
  • Advanced German engineering
  • Consistent performance in large-scale applications
  • Long-term durability and safety

Memmert UF vs UFplus Series – Key Differences

Feature Memmert UF Series Memmert UFplus Series
Control System SingleDISPLAY ControlCOCKPIT TwinDISPLAY ControlCOCKPIT
Display 1 TFT display 2 TFT displays
Operation Basic touch control Advanced dual-screen control
Temperature Sensors 1 Pt100 sensor 2 Pt100 sensors (redundant monitoring)
Monitoring Standard Dual monitoring (higher safety)
Heating Control Standard HeatBALANCE (adjustable top/bottom heating)
Programmability Basic Advanced programming & process control
Data Logging External / optional Built-in logging + display + USB export
Data Access Ethernet Ethernet + USB + onboard display logs
User Functions Limited Extended (User-ID, logs, program control)
Safety System Basic overtemperature protection AutoSAFETY + multi-level protection
Connectivity Ethernet Ethernet + USB + program transfer
Application Level Routine lab work Advanced research & controlled processes

Why Work With Copens Scientific?

  • Genuine Memmert laboratory equipment
  • Expert technical consultation for your applications
  • Reliable after-sales support and assistance
  • Experience supporting laboratories across Malaysia
  • Responsive service for enquiries and quotations

Get Expert Advice or Custom Solutions?

On this page, you’ll find the essential technical details of the Memmert universal oven.

If you need further clarification, our customer support team is ready to assist with product selection, specifications, and application suitability.

For customised solutions or special requirements, feel free to contact our technical specialists at info@copens-sci.com

Technical Specification

Temperature

Working temperature range at least 5 (UN/UNplus/UNm/UNmplus) or 10 (UF/UFplus/UFm/UFmplus) above ambient temperature to +300 °C
Setting accuracy temperature up to 99.9 °C: 0.1 / from 100 °C: 0.5
Setting temperature range +20 to +300 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT the process time does not start until the set temperature is reached
Calibration three freely selectable temperature values
adjustable parameters temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime

Ventilation

Fan forced air circulation by 2 quiet air turbines, adjustable in 10 % steps for each segment individually
Fresh air Admixture of pre-heated fresh air by electronically adjustable air flap
Vent vent connection with restrictor flap

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature
Temperature control overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
Autodiagnostic system for fault analysis
Alarm visual and acoustic

Standard equipment

Works calibration certificate Calibration at +160°C
Door fully insulated stainless steel doors with
2-point locking (compression door lock)
Internals 2 stainless steel grid(s), electropolished

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 1040 x 1200 x 600 mm (d less 39 mm for fan)
Interior easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
Volume 749 l
Max. number of internals 14
Max. loading of chamber 300 kg
Max. loading per internal 30 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 1224 x 1720 x 784 mm (d +56mm door handle)
Installation on lockable castors
Housing rear zinc-plated steel

Electrical data

Voltage
Electrical load
400 V and 3x 230 V w/o neutral, 50/60 Hz approx. 7000 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Altitude of installation max. 2,000 m above sea level
Ambient temperature +5 °C to +40 °C
Humidity rh max. 80 %, non-condensing
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 1330 x 1910 x 1050 mm
Net weight approx. 217 kg
Gross weight carton approx. 288 kg

Standard units are safety-approved and bear the test marks

CE Logo
VDE GS
UL
EAC Logo
×