Memmert ICH260L Climate Chamber

The Memmert ICH260L Climate Chamber is designed for stability testing where reliable temperature and humidity control are required with a streamlined feature set. Its 256-litre chamber provides sufficient capacity for routine testing while maintaining consistent environmental conditions.

This model supports stability testing in accordance with international guidelines such as ICH, WHO, EMA, ASEAN, GMP, GLP, and GCCP. It is suitable for laboratories that require dependable performance without the need for advanced configuration options.

Category: Tag: Brand:

Memmert ICH260L Climate Chamber

Key Features of Memmert ICH260L

  • 256-litre chamber for routine stability testing
  • Compressor-based cooling for stable temperature control
  • Precise temperature and humidity regulation
  • Designed for compliance with stability testing standards
  • Simplified configuration for straightforward operation
  • Stainless steel interior for durability
  • Reliable performance for laboratory environments

Climate Control Performance (ICH Series)

Performance depends on factors such as sample load, ambient conditions, and operating parameters. The ICH series is designed to maintain stable conditions even when external variables change.

Its compressor-based system allows consistent cooling and heating across a wider range, making it suitable for stability studies that require controlled variation and repeatability.

Temperature & Humidity Working Range (ICH Series)

Climate chambers can perform very differently depending on operating conditions. Factors such as the moisture content of the samples, surrounding ambient environment, and the selected temperature–humidity range all influence overall performance.

The ICH series is designed to support a wide combination of temperature and humidity settings. The working range diagram illustrates how these parameters can be adjusted to meet different testing requirements, including stability studies under regulated conditions.

Within the specified operating range, the chamber can maintain continuous, condensation-free operation under standard conditions. Near the boundary limits, condensation may occur depending on the sample load and ambient humidity.

Applications

The Memmert ICH260L climate chamber is suitable for:

  • Routine stability testing under ICH guidelines
  • Controlled storage and shelf-life studies
  • Quality control and validation processes
  • Environmental simulation for standard applications
  • Laboratory testing requiring reliable conditions

Compliance & Stability Standards

ICH260L is designed to support testing in accordance with internationally recognised standards, including:

  • ICH (International Council for Harmonisation)
  • WHO (World Health Organization)
  • EMA (European Medicines Agency)
  • ASEAN stability guidelines
  • GMP (Good Manufacturing Practice)
  • GLP (Good Laboratory Practice)
  • GCCP (Good Clinical Practice)

These standards require strict control of environmental conditions during stability studies. The chamber provides the consistency needed to meet these requirements.

Why Choose Memmert Equipment?

Memmert systems are known for:

  • Accurate environmental control
  • Reliable performance in regulated environments
  • High-quality German engineering
  • Long-term durability

Climate Chamber vs Humidity Chamber vs Constant Climate Chamber

Type What It Does Ideal For
Climate Chamber Controls temperature and humidity over a wide operating range (heating + cooling) Environmental simulation, stability testing, pharmaceutical and lab research
Humidity Chamber Focuses on high humidity conditions with temperature support 85/85 testing, accelerated life testing, moisture and electronics reliability testing
Constant Climate Chamber Maintains highly stable temperature and humidity with minimal fluctuation Stability testing, long-term storage, pharmaceutical and ICH-compliant studies

Why Work With Copens Scientific?

  • Genuine Memmert laboratory equipment
  • Expert technical consultation for your applications
  • Reliable after-sales support and assistance
  • Experience supporting laboratories across Malaysia
  • Responsive service for enquiries and quotations

Get Expert Advice or Custom Solutions?

On this page, you’ll find the essential technical details of the Memmert climate chamber.

If you need further clarification, our customer support team is ready to assist with product selection, specifications, and application suitability.

For customised solutions or special requirements, feel free to contact our technical specialists at info@copens-sci.com

Technical Specification

Temperature

Working temperature range without humidity: 0 to +60 °C
Setting temperature range 0 to +60 °C
Working temperature range from +10°C up to +60°C with humidity and/or light
Setting accuracy temperature 0.1 °C
Temperature sensor 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error

Humidity

Humidity humidity supply with distilled water from external tank by self-priming pump
Humidification humidification by hot steam generator
Humidity active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 %

Control of standard components

Light daylight and UV-light can be programmed individually via controller
Light cold white light +/- 6 %, for UV-light +/- 10 %
Light illumination unit consisting of 4 fluorescent lights with cold white light (standard illuminant D65, 6.500K) and 2 UV-lamps in the spectral range of 320-400 nm

Control technology

ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT the process time does not start until the set temperature is reached
adjustable parameters temperature (Celsius or Fahrenheit), fan speed, light, programme time, time zones, summertime/wintertime

Ventilation

Fan forced air circulation in working chamber adjustable from 10 to 100 % in 10 % steps

Communication

Documentation programme stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
Temperature control over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control
Alarm visual and acoustic

Heating concept

Air jacket air jacket heating system for gentle all-around heating
Heating concept no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior
Defrosting highly efficient automatic defrosting system

Standard equipment

Works calibration certificate standard values +10 °C and +37 °C, 60 % rh at +30 °C
Entry port Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed
by a silicone stopper, standard position at the back
Internals 2 stainless steel grid(s), electropolished
Standard accessories Water tank including connection hose

Stainless steel interior

Dimensions w(A) x h(B) x d(C): 640 x 800 x 500 mm (d less 33 mm for fan)
Volume 256 l
Max. number of internals 9
Max. loading of chamber 200 kg
Max. loading per internal 20 kg

Textured stainless steel casing

Dimensions w(D) x h(E) x d(F): 824 x 1552 x 685 mm (d +56mm door handle)
Installation on lockable castors
Housing rear zinc-plated steel

Electrical data

Voltage
Electrical load
230 V, 50/60 Hz
approx. 1450 W

Ambient conditions

Set Up The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature 15 °C to 28 °C (up to 34 °C with limited temperature
range)
Humidity rh max. 70 %, non-condensing
Altitude of installation max. 2,000 m above sea level
Overvoltage category II
Pollution degree 2

Packing/shipping data

Transport information The appliances must be transported upright
Customs tariff number 8419 8998
Country of origin Federal Republic of Germany
WEEE-Reg.-No. DE 66812464
Dimensions approx incl. carton w x h x d: 930 x 1760 x 930 mm
Net weight approx. 160 kg
Gross weight carton approx. 217 kg

Standard units are safety-approved and bear the test marks

CE Logo
EAC Logo
×
×