The Memmert HPP110eco Peltier Constant Climate Chamber is designed for reliable temperature and humidity control in laboratory environments. It uses energy-efficient Peltier technology to maintain stable conditions, making it suitable for long-term testing and sensitive applications.
Laboratories often rely on consistent environmental control for accurate results. This model provides a controlled climate with minimal energy consumption, supporting applications such as stability testing, storage, and research.
The HPPeco series uses advanced Peltier technology to provide both heating and cooling within a single system. Instead of relying on compressor-based refrigeration, it transfers heat directly using thermoelectric elements.
This design allows precise temperature control while maintaining stable environmental conditions. It also operates quietly and without refrigerants, making it a more energy-efficient and environmentally friendly solution for laboratory use.
The HPPeco climate chamber operates from temperatures starting as low as 0°C, providing a wide working range suitable for various laboratory applications. This flexibility allows precise control of both temperature and humidity under different testing conditions.
Designed for stability studies in accordance with ICH guidelines (Q1A), the chamber supports testing requirements in pharmaceutical, cosmetic, and food industries. It is also well suited for environmental simulation and material testing, where consistent conditions are essential.
With its ability to maintain stable environments across a broad temperature and humidity range, the system ensures reliable and repeatable results for long-term studies.
The Memmert HPP110eco peltier constant climate chamber is suitable for:
Memmert systems are trusted for:
| Type | What It Does | Ideal For |
|---|---|---|
| Climate Chamber | Controls temperature and humidity over a wide operating range (heating + cooling) | Environmental simulation, stability testing, pharmaceutical and lab research |
| Humidity Chamber | Focuses on high humidity conditions with temperature support | 85/85 testing, accelerated life testing, moisture and electronics reliability testing |
| Constant Climate Chamber | Maintains highly stable temperature and humidity with minimal fluctuation | Stability testing, long-term storage, pharmaceutical and ICH-compliant studies |
On this page, you’ll find the essential technical details of the Memmert universal oven.
If you need further clarification, our customer support team is ready to assist with product selection, specifications, and application suitability.
For customised solutions or special requirements, feel free to contact our technical specialists at info@copens-sci.com
| Setting temperature range | without light, with humidity: +5 to +70 °C |
|---|---|
| Setting temperature range | with light, with humidity: +5 to +70 °C |
| Setting temperature range | with light, without humidity: 0 to +70 °C |
| Setting temperature range | without light, without humidity: 0 to +70 °C |
| Working temperature range | with light, without or with humidity: 15 to +40 °C |
| Working temperature range | without light, with humidity: +5 (at least 20 below ambient temperature) to +70 °C |
| Working temperature range | without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C |
| Setting accuracy temperature | 0.1 °C |
| Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
| Setting range humidity | 10 - 90 % rh |
|---|---|
| Humidity | humidity supply with distilled water from external tank by self-priming pump |
| Humidification | humidification by hot steam generator |
| Dehumidification | dehumidification by cold trap using the Peltier technology |
| Setting accuracy humidity | 0.5 % rh |
| ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
|---|---|
| Language setting | German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese |
| adjustable parameters | temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime |
| Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
| Function SetpointWAIT | the process time does not start until the set temperature is reached |
| Calibration | three freely selectable values each, temperature and humidity |
| Interface | Ethernet LAN, USB |
|---|---|
| Documentation | programme stored in case of power failure |
| Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
| Temperature control | over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display |
|---|---|
| AutoSAFETY | additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature |
| Autodiagnostic system | integral fault diagnostics for temperature and humidity control |
| Alarm | visual and acoustic |
| Peltier | energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle) |
|---|
| Standard works calibration certificate | +25 °C / 60 % rh, +40 °C / 75 % rh, 5 % CO2 at +40 °C / 75 %rh |
|---|---|
| Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
| Door | inner glass door |
| Entry Port | Entry port (silicone), 18 mm clear diameter, moisture tight, can be closed by silicone stopper, in the side panel, centre left |
| Internals | 2 stainless steel grid(s), electropolished |
| Standard accessories | Water tank including connection hose |
| Dimensions | w(A) x h(B) x d(C): 560 x 480 x 400 mm (d less 32 mm for fan - Peltier) |
|---|---|
| Volume | 108 l |
| Max. number of internals | 5 |
| Max. loading of chamber | 150 kg |
| Max. loading per internal | 20 kg |
| Dimensions | w(D) x h(E) x d(F): 745 x 864 x 555 mm (d +56mm door handle & +111mm Peltier element) |
|---|---|
| Housing | rear zinc-plated steel |
| Voltage Electrical load |
230 V, 50/60 Hz approx. 420 W |
|---|
| Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
|---|---|
| Ambient temperature | 16 ºC to 40 ºC |
| Humidity rh | max. 70 %, non-condensing |
| Altitude of installation | max. 2,000 m above sea level |
| Overvoltage category | II |
| Pollution degree | 2 |
| Country of origin | Federal Republic of Germany |
|---|---|
| Transport information | The appliances must be transported upright |
| Customs tariff number | 8419 8998 |
| WEEE-Reg.-No. | DE 66812464 |
| Dimensions approx incl. carton | w x h x d: 830 x 1050 x 800 mm |
| Net weight | approx. 64 kg |
| Gross weight carton | approx. 87 kg |