The Memmert ICH110 Climate Chamber is designed for stability testing where precise temperature and humidity control are required under regulated conditions. Its 108-litre chamber, combined with compressor-based cooling, allows reliable performance across a wide operating range.
This model is commonly used for stability testing in accordance with international guidelines such as ICH, WHO, EMA, ASEAN, GMP, GLP, and GCCP. It supports laboratories that need consistent and traceable environmental conditions for pharmaceutical, life science, and quality control applications.
Performance depends on factors such as sample load, ambient conditions, and operating parameters. The ICH series is designed to maintain stable conditions even when external variables change.
Its compressor-based system allows consistent cooling and heating across a wider range, making it suitable for stability studies that require controlled variation and repeatability.
Climate chambers can perform very differently depending on operating conditions. Factors such as the moisture content of the samples, surrounding ambient environment, and the selected temperature–humidity range all influence overall performance.
The ICH series is designed to support a wide combination of temperature and humidity settings. The working range diagram illustrates how these parameters can be adjusted to meet different testing requirements, including stability studies under regulated conditions.
Within the specified operating range, the chamber can maintain continuous, condensation-free operation under standard conditions. Near the boundary limits, condensation may occur depending on the sample load and ambient humidity.
The Memmert ICH110 climate chamber is suitable for:
ICH110 is designed to support testing in accordance with internationally recognised standards, including:
These standards require strict control of environmental conditions during stability studies. The chamber provides the consistency needed to meet these requirements.
Memmert systems are known for:
| Type | What It Does | Ideal For |
|---|---|---|
| Climate Chamber | Controls temperature and humidity over a wide operating range (heating + cooling) | Environmental simulation, stability testing, pharmaceutical and lab research |
| Humidity Chamber | Focuses on high humidity conditions with temperature support | 85/85 testing, accelerated life testing, moisture and electronics reliability testing |
| Constant Climate Chamber | Maintains highly stable temperature and humidity with minimal fluctuation | Stability testing, long-term storage, pharmaceutical and ICH-compliant studies |
On this page, you’ll find the essential technical details of the Memmert climate chamber.
If you need further clarification, our customer support team is ready to assist with product selection, specifications, and application suitability.
For customised solutions or special requirements, feel free to contact our technical specialists at info@copens-sci.com
| Setting temperature range | -10 to +60 °C |
|---|---|
| Working temperature range | from -10°C up to +60°C without humidity (not suitable for long-term storing at sub-zero temperatures. During permanent operation, the glass door may ice over) |
| Working temperature range | from +10°C up to +60°C with humidity and/or light |
| Setting accuracy temperature | 0.1 °C |
| Temperature sensor | 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
| Humidity | humidity supply with distilled water from external tank by self-priming pump |
|---|---|
| Humidification | humidification by hot steam generator |
| Humidity | active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 % |
| ControlCOCKPIT | TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
|---|---|
| adjustable parameters | temperature (Celsius or Fahrenheit), fan speed, relative humidity, programme time, time zones, summertime/wintertime |
| Language setting | German, English, Spanish, French, Polish, Czech, Hungarian |
| Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
| Function SetpointWAIT | the process time does not start until the set temperature is reached |
| Calibration | three freely selectable values each, temperature/humidity/CO2 |
| Documentation | programme stored in case of power failure |
|---|---|
| Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
| Temperature control | mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature |
|---|---|
| Temperature control | over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display |
| AutoSAFETY | additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature |
| Autodiagnostic system | integral fault diagnostics for temperature and humidity control |
| Alarm | visual and acoustic |
| Air jacket | air jacket heating system for gentle all-around heating |
|---|---|
| Heating concept | no drying-up of the load caused by the cooling device due to separation of thermal jacket from interior |
| Cooling | energy-saving use of CFC-free cooling/heating system (refrigerant R134a) |
| Defrosting | highly efficient automatic defrosting system |
| Entry port | Entry port (silicone), 40 mm clear diameter, moisture tight, can be closed by a silicone stopper, standard position at the back |
|---|---|
| Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
| Internals | 2 stainless steel grid(s), electropolished |
| Works calibration certificate | standard values +10 °C and +37 °C, 60 % rh at +30 °C |
| Door | inner glass door |
| Dimensions | w(A) x h(B) x d(C): 560 x 480 x 400 mm (d less 33 mm for fan) |
|---|---|
| Volume | 108 l |
| Max. number of internals | 5 |
| Max. loading of chamber | 150 kg |
| Max. loading per internal | 20 kg |
| Dimensions | w(D) x h(E) x d(F): 745 x 1233 x 585 mm (d +56mm door handle) |
|---|---|
| Installation | on lockable castors |
| Housing | rear zinc-plated steel |
| Voltage Electrical load | 230 V, 50/60 Hz approx. 1350 W |
|---|
| Set Up | The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
|---|---|
| Ambient temperature | 15 °C to 28 °C (up to 34 °C with limited temperature range) |
| Humidity rh | max. 70 %, non-condensing |
| Altitude of installation | max. 2,000 m above sea level |
| Overvoltage category | II |
| Pollution degree | 2 |
| Transport information | The appliances must be transported upright |
|---|---|
| Customs tariff number | 8419 8998 |
| Country of origin | Federal Republic of Germany |
| WEEE-Reg.-No. | DE 66812464 |
| Dimensions approx incl. carton | w x h x d: 880 x 1410 x 810 mm |
| Net weight | approx. 109 kg |
| Gross weight carton | approx. 137 kg |